16. November 2021
Can the unit select the drilling depth depending on the tilt angle?
Press the red button for a long time > Tilt compensation: ON Now the tilt compensation is switched on. If the drilling depth preselection is switched on, the machine will select the actual drilling depth according to the set diameter/radius…
16. November 2021
Is it possible to adjust the angle of inclination of the unit to the position of the object to be tested?
Press the device against the object to be tested. When the safety button is pressed, press the red and black button simultaneously. The tilt indicator in the display now lights up green and the REL symbol appears in the upper…
16. November 2021
When is a drilling needle worn out?
The needle wear depends on several factors. You will find an illustration of this in the maintenance instructions (page 14) enclosed with your machine.
16. November 2021
When is a drill chuck defective?
A drill chuck should be replaced when irregular resistance is felt when turning by hand, or the rear “offset value” is over 400. (see illustration)
16. November 2021
The display of the unit rotates when the position is changed. Can this process be switched off?
Control Panel > General > Rotate Display: NO Now the rotation mode of the display is switched off.
16. November 2021
A Bluetooth connection between the device and the computer is no longer possible. What could be the reason for this?
The Bluetooth must be turned on. And the correct mode must be selected for your device pairing. If there is still no connection, try out to update your Bluetooth driver. Instructions can be found under Instruction manuals & software downloads in the…
16. November 2021
What is the difference between passive and active Bluetooth pairing?
– Active Bluetooth mode means the pairing is started by the drill. You must create a new connection in order to pair. To do this, you must find the computer on the side of the device and pair it. –…
16. November 2021
Why does the needle drills into a cavity and automatically retracts at the end of the cavity?
This activates the automatic needle retraction. This is a protective mechanism that counteracts excessive bending of the needle. It is set to 5 cm by default. The retraction length can be adjusted under Control Panel > Automatic Needle Retraction. !…
15. November 2021
The automatic needle check always displays “Change needle” even if the needle is new. Why are new needles shown as worn out?
The test disk may be worn out. The test disc is a replaceable disc in the front adapter. If the needle is pulled through the test disc during the needle test, the needle is rated as bad. It is possible…